Scan View SCANMATE F10 Instrukcja Obsługi Strona 41

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 40
Adjusting the alignment of the transmission lamp
1 • If the self-test shows that the transmission lamp is not properly aligned,
insert the special plastic-covered screwdriver (supplied with the scanner)
in the space between the scanner lid and the lamp shield to move the
lamp assembly
very slightly
(ca. 1 mm) in the indicated direction:
front end to the right = apply screwdriver at position 1
rear end to the right = apply screwdriver at position 2
front end to the left = apply screwdriver at position 3
rear end to the left = apply screwdriver at position 4
CAUTION: Use the screwdriver as a lever. Do not twist it.
2 • After making the adjustment, click on “Continue” to repeat the
alignment test.
3 • If correct alignment cannot be achieved after several tries, please
contact your ScanView dealer for assistance.
Maintenance
ScanMate F10 • Operator’s Manual
6•5
1
2
3
4
If adjustment is necessary, use the flat side of
the screwdriver as shown to move the lamp
assembly very slightly in the required direction.
Do not twist the screwdriver.
Przeglądanie stron 40
1 2 ... 36 37 38 39 40 41 42 43 44 45 46 ... 49 50

Komentarze do niniejszej Instrukcji

Brak uwag